TEMPERATURE MEASUREMENT
Scale Range: -40°C to 1100°C or -40°F to 2000°F;
Models with Thermocouple Type J or K Outputs:
-30
°C to 760°C or -30°F to 1390°
F
Resolution: 1.0°C or 1.0°F
Accuracy: ±1.0% of Reading, ±1.0°C or °F
Repeatability: ±1.0°C or ±2.0°F
Temperature: All Functions in °C or °F
Noise Effective Temperature: <1.0°C
OPTICAL CONSIDERATIONS
Optical Configuration: Robust, Aerospace-Quality, Double-Coated Zinc Selenide Optics per Military Specification MIL-C-13508
Spectral Pass Band: 8 < Wavelengths < 14 Microns
Sighting: Line of Sight
Field of View: Standard:  2° or 15°
OPERATING CONDITIONS
Operating Environment: -10°C to 70°C or 14°F to 160°F, up to 99% Relative Humidity, Non-Condensing
Storage Temperature: Same as Operating Environment Temperature
ELECTRICAL INTERFACE
Power Requirements: 5V to 26V DC (Current Draw: 10 mA + Current Signal)
Output Signal: Standard: Analog mV (1.0mV/°C)
Optional: 4-20 mA, 0-5 Volt and Type J or K Simulated Thermocouple
GENERAL
Response Time: 0.02 Second
Emissivity Compensation: Preset at 0.98 or at End-User’s Request
Operating Distance: 2 cm to 300 meters
Dimensions: 2" Diameter by 3" Length
Mounting: Ball & Socket Swivel Mount
Warranty: One Year Limited Warranty on Parts & Labor

Everest Interscience, Inc. has a policy of continued design upgrade. Therefore, we reserve the right to change specifications without notice.